Follow-up of intracranial aneurysms treated by flow diverter: comparison of three-dimensional time-of-flight MR angiography (3D-TOF-MRA) and contrast-enhanced MR angiography (CE-MRA) sequences with digital subtraction angiography as the gold standard
Abstract
Background and purpose: Follow-up of intracranial aneurysms treated by flow diverter with MRI is complicated by imaging artifacts produced by these devices. This study compares the diagnostic accuracy of three-dimensional time-of-flight MR angiography (3D-TOF-MRA) and contrast-enhanced MRA (CE-MRA) at 3 T for the evaluation of aneurysm occlusion and parent artery patency after flow diversion treatment, with digital subtraction angiography (DSA) as the gold standard.
Materials and methods: Patients treated with flow diverters between January 2009 and January 2013 followed by MRA at 3 T (3D-TOF-MRA and CE-MRA) and DSA within a 48 h period were included in a prospective single-center study. Aneurysm occlusion was assessed with full and simplified Montreal scales and parent artery patency with three-grade and two-grade scales.
Results: Twenty-two patients harboring 23 treated aneurysms were included. Interobserver agreement using simplified scales for occlusion (Montreal) and parent artery patency were higher for DSA (0.88 and 0.61) and CE-MRA (0.74 and 0.55) than for 3D-TOF-MRA (0.51 and 0.02). Intermodality agreement was higher for CE-MRA (0.88 and 0.32) than for 3D-TOF-MRA (0.59 and 0.11). CE-MRA yielded better accuracy than 3D-TOF-MRA for aneurysm remnant detection (sensitivity 83% vs 50%; specificity 100% vs 100%) and for the status of the parent artery (specificity 63% vs 32%; sensitivity 100% vs 100%).
Conclusions: At 3 T, CE-MRA is superior to 3D-TOF-MRA for the evaluation of aneurysm occlusion and parent artery patency after flow diversion treatment. However, intraluminal evaluation remains difficult with MRA regardless of the sequence used.